Description
Spectrophotometer FUV-1800/FUV-1600/FV-1800/FV-1600
| Model | FUV-1800 FUV-1800PC |
FUV-1600 FUV1600PC |
FV-1800 FV-1800PC |
FV-1600 FV-1600PC |
| Optical System | Single Beam,Grating 1200 lines/mm | |||
| Wavelength Range |
190-1100nm | 320-1100nm | ||
| Spectral Bandwidth |
2nm | 4nm | 2nm | 4nm |
| Wavelength Accuracy |
±0.5nm | |||
| Wavelength Repeatability |
≤0.2nm | |||
| Photometric Accuracy |
±0.002A(0-0.Abs), ±0.004A(0.5-1.0Abs), ±0.5%T(0-100%T) |
|||
| Photometric Repeatability |
0.001Abs(0-0.5Abs), 0.002Abs(0.5-1.0Abs), ≤0.2%T (0-100%T) |
|||
| Stray Light | ≤0.04%T@360nm;220nm | ≤0.04%T@360nm | ||
| Stability | ±0.001A/h@500nm | |||
| Baseline Flatness | ±0.002A | |||
| Noise | 0.0005A | |||
| Display | 65 thousand ture color TFT LCD (480*272) | |||
| Photometric Mode |
T,A,C,E | |||
| Photometric Range |
0-200% T,-0.301-3.0A | |||
| Detector | Silicon Photodiode | |||
| Light Source |
Deuterium Lamp,Tungsten Lamp | |||
| Input | Membrane Keypad | |||
| Output | USB-A*2Print and data output USB-B Connect PC | |||
| Compartment | Optional 8Auto Cell holder, Solid Sample Holder, Micro Cell Holder, 100mm Cell Holder |
|||